Late adverse event reporting from medical device manufacturers to the US Food and Drug Administration: cross sectional study

BMJ 2025; 388 doi: https://doi.org/10.1136/bmj-2024-081518 (Published 12 March 2025) Cite this as: BMJ 2025;388:e081518

Linked Editorial

Active postmarketing device surveillance as a legislative priority

  1. Alexander O Everhart, instructor1 2,
  2. Pinar Karaca-Mandic, professor3,
  3. Rita F Redberg, professor4 5,
  4. Joseph S Ross, professor6 7 8,
  5. Sanket S Dhruva, associate professor9 10 11
  1. 1Division of General Medical Sciences, John T Milliken Department of Medicine, Washington University School of Medicine in St Louis, St Louis, MO, USA
  2. 2Center for Advancing Health Services, Policy and Economics Research, Washington University in St. Louis, St Louis, MO, USA
  3. 3Department of Finance, Carlson School of Management, University of Minnesota, Minneapolis, MN, USA
  4. 4Division of Cardiology, Department of Medicine, University of California, San Francisco School of Medicine, San Francisco, CA, USA
  5. 5Philip R. Lee Institute for Health Policy Studies, University of California, San Francisco School of Medicine, San Francisco, CA, USA
  6. 6Section of General Medicine and the National Clinician Scholars Program, Department of Internal Medicine, Yale School of Medicine, New Haven, CT, USA
  7. 7Department of Health Policy and Management, Yale School of Public Health, New Haven, CT, USA
  8. 8Center for Outcomes Research and Evaluation, Yale-New Haven Health System, New Haven, CT, USA
  9. 9Division of Cardiology, Department of Medicine, University of California, San Francisco School of Medicine, San Francisco, CA, USA
  10. 10Philip R Lee Institute for Health Policy Studies, University of California, San Francisco School of Medicine, San Francisco, CA, USA
  11. 11San Francisco Veterans Affairs Health Care System, San Francisco, CA, USA
  1. Correspondence to: A O Everhart everhart{at}wustl.edu (@aoeverhart.bsky.social on Bluesky)
  • Accepted 4 February 2025

Abstract

Objective To describe the extent of late adverse event reporting by manufacturers to the US Food and Drug Administration’s (FDA) Manufacturer And User Facility Device Experience (MAUDE) database as well as the distribution of late reporting among manufacturers and associations with device characteristics.

Design Cross sectional study.

Setting The FDA MAUDE database, a central postmarket safety surveillance tool for US medical devices, from 1 September 2019 to 31 December 2022.

Participants Medical device manufacturers that submitted initial adverse event reports to the FDA between 1 September 2019 and 31 December 2022.

Main outcome measures Time in days between date manufacturer was notified of event and date of FDA receipt of adverse event reports, proportion of reports reported late (after the required 30 day window as required by FDA regulation), and distribution of late reporting among manufacturers and medical devices.

Results 13 587 reports were of deaths, 1 552 268 of injuries, and 2 866 693 of malfunctions received by the FDA from 3028 unique manufacturers and 88 448 unique medical devices in the three and a half year period. Of 4 432 548 included reports, 71.0% (n=3 146 957) of adverse events were reported within 30 days (on time), 4.5% (n=197 606) were reported between 31 and 180 days (late), and 9.1% (n=402 891) were after 180 days (late). 15.5% of reports (n=685 094) had missing or invalid date data provided by the manufacturer. Three manufactures and 13 medical devices were attributed to 54.8% of late reports.

Conclusions Nearly a third of manufacturer reports of medical device adverse events were not demonstrably submitted to the FDA within the regulatory deadline, with most late reports being submitted more than six months after manufacturer notification. Most late reports were submitted by a small number of manufacturers. Late adverse event reporting may prevent early detection of patient safety concerns.

Introduction

Most medical devices in the US and globally receive regulatory authorization based on non-clinical data or clinical trials with limitations in their design.1234567891011 Given these limitations in premarket evidence, postmarket device safety surveillance is crucial. Postmarket surveillance is also increasingly important as the US Food and Drug Administration (FDA) explores and implements evaluations of existing postmarket data and future data collection plans during premarket authorization reviews. For example, through the Breakthrough Devices Program, the FDA may accept greater premarket uncertainty of the benefit-risk profile for highly novel (“breakthrough”) devices, provided that the manufacturer commits to sufficient postmarket data collection.12

A central postmarket safety surveillance tool for US medical devices is the FDA’s Manufacturer And User Facility Device Experience (MAUDE) database, where manufacturers and other stakeholders report device related adverse events and product problems (hereafter adverse events). MAUDE reports are the most common source of information used to initiate medical device safety communications, the FDA’s primary mechanism for communicating postmarket device safety issues.13

While a key tool in detecting safety issues, MAUDE has known limitations. MAUDE relies on external individuals and organizations to report adverse events, rather than systematically and prospectively collecting safety and outcome data. Moreover, reporting is at times inaccurate and incomplete; key clinical details can be omitted from reports, and reports describing patients “expiring” or “dying” are often categorized as injuries or malfunctions rather than deaths.141516

MAUDE reports may also not be timely. Over 95% of MAUDE reports are contributed by medical device manufacturers.17 Manufacturers are required by FDA regulation to submit adverse event reports to MAUDE within 30 days of becoming aware of them. However, recent media reports have described manufacturers withholding reports from MAUDE years beyond this deadline.1819202122 For example, the FDA determined that over the course of six years, a manufacturer of continuous positive airway pressure (CPAP) machines withheld hundreds of reports of injuries and deaths related to foam degradation in their machines, likely leading to otherwise preventable patient harm.18 These reports have resulted in a Government Accountability Office inquiry into the FDA’s oversight of manufacturer reporting and management of recalls.23 Studies have also found that both pharmaceutical and device manufacturers sometimes withhold safety information from the public.24252627 Delays in reporting may preclude regulators, clinicians, and patients from learning about emerging safety issues, resulting in avoidable patient harm.

This study quantified the timeliness of manufacturer reporting in the MAUDE database, including the distribution of late reporting among manufacturers and associations with device characteristics.

Methods

Medical device report data

This cross-sectional analysis was performed using the MAUDE database, a public database of adverse event reports involving medical devices.14 Reports include device names; generic device functions; manufacturers; whether events were deaths, injuries, or product malfunctions; and dates that the FDA received the reports. Reports of death identify instances where medical device use may have caused or contributed to patient deaths, while reports of injury are instances where device use may have caused or contributed to an injury or illness that is life threatening, resulted “in permanent impairment of a body function or permanent damage to a body structure,” or required a medical intervention to preclude permanent impairment or permanent damage.2829 Malfunctions are instances where devices did not meet performance specifications or otherwise perform as intended and did not cause or contribute to a death or injury but would likely contribute to a death or injury if the malfunction recurred.2829

Manufacturers, importers, and device user facilities (certain health care facilities) must report adverse events. Others, including clinicians and patients, may voluntarily report. Over 95% of MAUDE reports are made by manufacturers,17 who must submit reports within 30 days of becoming aware of events; manufacturers are required to report the dates that they were notified of events. Manufacturers are also required to submit reports within five working days for certain high risk events that require “requires remedial action to prevent an unreasonable risk of substantial harm to the public health,” although MAUDE does not indicate which reports were subject to this requirement.3031

For this study, manufacturer names as identified in MAUDE (or in device authorization databases if names were unavailable in MAUDE) were standardized by removing punctuation and division identifiers (eg, “Company X Neurology” and “COMPANY X, INC.” were standardized to “COMPANY X”).32 Individual devices were identified based on unique combinations of branded device name, generic device function, and manufacturer name.

Study sample

Our study sample included manufacturer reports received by the FDA between 1 September 2019 and 31 December 2022. Before 1 September 2019, manufacturers could submit summary reports that aggregated multiple reports into a single submission outside of MAUDE (summary reports within MAUDE are still permitted).2933 We limited the sample to initial reports, meaning records wherein manufacturers first learned of events (versus follow up reports that describe previous events). We excluded reports listed as summary reports, missing manufacturer names, with “Other” as the event type, or missing device class.

Medical device characteristics

We identified characteristics wherein differences in reporting times may inform policy making and clinical practice. These included characteristics determined by the time of FDA authorization (hereafter preauthorization characteristics), including whether devices were lifesaving or life sustaining (hereafter lifesaving), whether devices were implantable, whether devices qualified for the Breakthrough Devices Program, clinical specialty of device, and device class. Devices are assigned to one of three classes based on patient risk, which informs the evidence needed to evaluate their safety and effectiveness.34 Other characteristics included whether devices were the subject of an ongoing recall at the time that the FDA received the report or the subject of a terminated recall (ie, necessary corrective and preventive actions completed) when the FDA received the report.

Clinical specialty, class, lifesaving status, and implantable status were identified based on device product type codes and generic descriptions of device functions.35 Breakthrough status and recall status were assessed using public FDA data.3637 We considered all classes of recalls. Recall status was only assessed for class II and III (intermediate and high risk) devices3839; class I (low risk) devices do not have individual authorization records and cannot be reliably linked to FDA recall data.

Outcomes

The main outcome of interest was report time, the difference between the date a manufacturer reported that they were notified of an event and the date the FDA received the report. Report times were classified as filed in 0 to 30 days (required by regulation), 31 to 180 days, 181 days or greater, less than 0 days (reporting error caused by an invalid date), or missing manufacturer notification date (missing report times). Report times greater than 30 days were considered late.

Statistical analysis

Means and percentiles for non-negative and non-missing report times were calculated. Proportions of reports that were reported late or had missing or invalid notification dates were calculated for each manufacturer and medical device. To characterize concentration in reporting patterns, manufacturers and devices were ranked on the basis of numbers of late reports, on time reports, and reports with missing or invalid dates. Distributions of late and on time reports across months were calculated to assess whether reports were released consistently throughout the sample period or concentrated in certain months; concentration in months was assessed via Gini coefficients.40

Z tests were performed to compare differences in Gini coefficients and in late reporting by device characteristics among reports with non-negative and non-missing report times. Cuzick tests were performed to assess whether month-by-month trends were present in late reporting.41 Q values were calculated via the Simes procedure to adjust two sided P values for multiple testing. A significance threshold of Q<0.05 was used, implying a false discovery rate of 5%.424344 Statistical analyses were performed using Stata 15 SE.

Patient and public involvement

We did not directly involve patients or the public as part of the design or conduct of this study because we did not have funding to do so. However, this study was motivated in part by media reports of patient harm potentially caused by late reporting.1819202122 Additionally, as we prepared our findings for publication, we reviewed results with the leader of a patient and stakeholder advocacy group to ensure the importance of the work and revised the manuscript to improve its clarity, particularly results reporting and the implications of our findings.

Results

Study sample

Manufacturers reported 4 528 153 initial medical device reports in the MAUDE database to the FDA between 1 September 2019 and 31 December 2022. We excluded 57 779 (1.3%) for being a summary report and 37 826 (0.8%) for missing manufacturer name, device class, or event type. The resulting sample included 4 432 548 reports (98% of all initial manufacturer reports) (appendix table 1). These included 13 587 deaths, 1 552 268 injuries, and 2 866 693 malfunctions from 3028 unique manufacturers and 88 448 unique devices (table 1).

Table 1

Medical device report times by event type

CharacteristicDeath (n=13 587)Injury (n=1 552 268)Product malfunction (n=2 866 693)Total (n=4 432 548)
Unique manufacturers397194920323028
Unique devices258448 00349 37288 448
Report time, mean (SD)*†54.3 (171.1)61.7 (151.1)99.5 (270.2)89.6 (245.2)
Report time, median (IQR)*†23 (13)22 (15)17 (20)19 (20)
Reported within 30 days, No. (%)12 408 (91.3)772 543 (49.8)2 362 006 (82.4)3 146 957 (71.0)
Reported from 31 to 180 days, No. (%)377 (2.8)126 431 (8.1)70 798 (2.5)197 606 (4.5)
Reported over 180 days, No. (%)627 (4.6)71 620 (4.6)330 644 (11.5)402 891 (9.1)
Missing or invalid date, No. (%)175 (1.3)581 674 (37.5)103 245 (3.6)685 094 (15.5)
  • SD=standard deviation; IQR=interquartile range.

  • * Report time defined based on difference between day Food and Drug Administration reported receiving device report and day manufacturer reported becoming aware of device event.

  • Mean report time and percentiles only calculated for non-missing and non-negative report times.

Table 1

Medical device report times by event type

Extent of late reporting

Overall, 3 146 957 (71.0%) reports were made within 30 days (on time), 197 606 (4.5%) were between 31 and 180 days (late), 402 891 (9.1%) were after 180 days (late), and 685 094 (15.5%) had missing or invalid receipt dates (dates suggesting negative report times) (table 1). Among the 13.6% of reports made late, 66.9% were reported after 180 days. 1004 deaths (7.4%) were reported late, as were 198 051 (12.7%) injuries and 401 442 (14.0%) malfunctions. On time reporting significantly declined over time both overall (Q=0.02) and for deaths (Q=0.02) and injuries (Q<0.001) specifically, while on time reporting for malfunctions was stable over time (Q=0.81) (appendix figures 1-4). Approximately 99% of reports with missing or invalid receipt dates were categorized as such due to missing receipt dates (appendix table 2).

On time reporting rates ranged from 24.9% for dental devices to 99.7% for immunology devices (appendix table 3). Appendix figure 5 presents histograms of late report times.

Distribution of late reporting

Late reports were concentrated among a few manufacturers: three (0.1%) manufacturers contributed 54.8% of late reports (fig 1). The mean late reporting rate at the manufacturer level was 23.4% (fig 2). Among the top 200 manufacturers as ranked by total reports, 68.5% of manufacturers had fewer than 10% of their reports submitted late. The top 10 manufacturers ranked by the highest total number of late reports were: Becton Dickinson, Medtronic, Dentsply, Abbott, Boston Scientific, Nobel Biocare, Dexcom, Altatec, Biohorizons Implant Systems, and Institut Straumann (table 2). Reports with missing or invalid dates were more concentrated among manufacturers, while on time reports were less concentrated. (appendix table 4 and appendix figures 7-8).

Fig 1
Fig 1

Cumulative proportion of medical device reports among top 200 manufacturers with most late reports. Report time, defined as the difference between the day the FDA received a device report and the day the manufacturer became aware of the event, was considered late if it exceeds 30 days. Top 200 manufacturers ranked on late reports accounted for over 99% of all late device reports. Created using the analytical sample presented in appendix figure 6

Fig 2
Fig 2

Histogram distribution of proportion of medical device reports reported late among top 200 manufacturers with most total reports. Report time, defined as the difference between the day the FDA received a device report and the day the manufacturer became aware of the event, was considered late if it exceeds 30 days. Top 200 manufacturers ranked on total reports accounted for over 99% of all device reports. Created using the analytical sample presented in appendix figure 6

Table 2

Ranking based on the highest number of late medical device reports by top 10 manufacturers

RankNameTotal reportsTotal late reportsPercent of all late reports, %
1Becton Dickinson1 087 292151 61725.2
2Medtronic415 809116 95819.5
3Dentsply117 48160 72710.1
4Abbott140 96457 8329.6
5Boston Scientific112 13236 7676.1
6Nobel Biocare366 44825 3624.2
7Dexcom542 33318 5673.1
8Altatec35 52918 1613.0
9Biohorizons Implant Systems42 56917 0972.8
10Institut Straumann316 85114 1682.4
Table 2

Ranking based on the highest number of late medical device reports by top 10 manufacturers

Late reports were similarly concentrated among a few devices: 13 (<0.1%) devices accounted for 50.4% of late reports (fig 3). Overall, 71.5% of the top 200 devices had fewer than 10% of their reports submitted late (fig 4). The top 10 devices as ranked by total late reports included Becton Dickinson’s infusion pumps, Abbott’s glucose monitors, Medtronic’s insulin pumps, Biohorizons Implant Systems’ dental implants, and Dexcom’s glucose monitors (table 3).

Fig 3
Fig 3

Cumulative proportion of medical device reports among top 200 devices with most late reports. Report time, defined as the difference between the day the FDA received a device report and the day the manufacturer became aware of the event, was considered late if it exceeds 30 days. Top 200 devices ranked on late reports accounted for over 78% of all late device reports. Created using the analytical sample presented in appendix figure 6

Fig 4
Fig 4

Histogram distribution of proportion of medical device reports reported late among top 200 devices with most total reports. Report time, defined as the difference between the day the FDA received a device report and the day the manufacturer became aware of the event, was considered late if it exceeds 30 days. Top 200 medical devices ranked on total reports accounted for over 67% of all device reports. Created using the analytical sample presented in appendix figure 6

Table 3

Ranking based on number of late medical device reports by top 10 devices

RankProduct nameGeneric nameManufacturer nameTotal reportsTotal late reportsPercent of all late reports, %
18100 Alaris pump modulePump, infusionBecton Dickinson139 140102 07917.0
2Libre sensor freestyleFlash glucose monitoring systemAbbott52 95842 3097.0
38015 Alaris system pc unitPump, infusionBecton Dickinson45 35731 5625.3
4Pump mmt-1780kpk 670 g pathway black mgArtificial pancreas device system, single hormonal controlMedtronic60 32426 8574.5
5Biohorizons dental implantDental implantBiohorizons Implant Systems42 49317 0402.8
6640 g insulin pump mmt-1712kPump, infusion, insulin, to be used with invasive glucose sensorMedtronic26 66215 1232.5
7Dexcom g6 continuous glucose monitoring systemContinuous glucose monitorDexcom512 37812 7592.1
8640 g insulin pump mmt-1711kPump, infusion, insulin, to be used with invasive glucose sensorMedtronic18 83712 6522.1
9630 g insulin pump mmt-1715k 630 g black mgArtificial pancreas device system, threshold suspendMedtronic21 62711 4031.9
108110 Alaris syringe pumpPump, infusionBecton Dickinson19 51599761.7
  • Manufacturers identified based on standardized version (removing extraneous punctuation, abbreviations, etc) of manufacturer name as reported in manufacturer and user facility device experience database or medical device approval or clearance databases if name was unavailable in the manufacturer and user facility device experience database. Medical devices are identified based on unique combinations of branded device name, generic device name, and manufacturer name. Report time defined based on difference between day Food and Drug Administration reported receiving device report and day manufacturer reported becoming aware of device event. Late reports are defined as those with report times greater than 30 days. “percent of all late reports” denotes what percent of all late medical device reports during the sample period were attributed to the given manufacturer or medical device.

Table 3

Ranking based on number of late medical device reports by top 10 devices

Late reports were more likely to be released in short time periods (Gini coefficient=0.554), while reports that were on time were more gradually released throughout the study period (Gini coefficient=0.278; Q<0.001) (appendix fig 9). Similar patterns were observed for death, injury, and malfunction reports (appendix fig 10-12). Distributions of reports over time for the top 10 devices as ranked by total late reports are presented in appendix figures 13-22.

Late reporting by preauthorization characteristics

Among reports with no missing or invalid report times, higher risk devices had higher percentages of late reports: 6.9% (95% confidence interval (CI) 6.7% to 7.1%) of the 70 160 reports for class I (low risk) devices were late compared with 13.8% (13.8% to 13.9%, Q<0.001) of the 2 667 628 reports for class II (intermediate risk) devices, and 22.4% (22.3% to 22.5%, Q<0.001) of the 1 009 666 reports for class III (high risk) devices (table 4). Late reporting by class varied by event type. Class III devices had the lowest percentage of injury reports reported late but the highest percentage of malfunction reports reported late. No significant differences were noted in late death reports by device class.

Table 4

Differences in percent reported late by event type and preauthorization device characteristics

Pre-authorization device characteristicsAllDeathInjuryMalfunction
Total No.Reported lateQTotal No.Reported lateQTotal No.Reported LateQTotal No.Reported lateQ
No.% (95% CI)No.% (95% CI)No.% (95% CI)No.% (95% CI)
Device class
Class I70 16048666.9 (6.7 to 7.1)Ref1041211.5 (5.4 to 17.7)Ref6360108517.1 (16.1 to 18.0)Ref63 69637695.9 (5.7 to 6.1)Ref
Class II2 667 628369 36913.8 (13.8 to 13.9)<0.00129942929.8 (8.7 to 10.8)0.59586 602153 23526.1 (26.0 to 26.2)<0.0012 078 032215 84210.4 (10.3 to 10.4)<0.001
Class III1 009 666226 26222.4 (22.3 to 22.5)<0.00110 3147006.8 (6.3 to 7.3)0.06377 63243 73111.6 (11.5 to 11.7)<0.001621 720181 83129.2 (29.1 to 29.4)<0.001
Lifesaving device
Non-life saving3 476 956557 05916.0 (16.0 to 16.1)Ref86227979.2 (8.6 to 9.9)Ref862 051173 78120.2 (20.1 to 20.2)Ref2 606 283382 48114.7 (14.6 to 14.7)Ref
Life saving270 49843 43816.1 (15.9 to 16.2)0.6447902074.3 (3.7 to 4.9)<0.001108 54324 27022.4 (22.1 to 22.6)<0.001157 16518 96112.1 (11.9 to 12.2)<0.001
Implantable device
Non-implantable2 778 561394 01314.2 (14.1 to 14.2)Ref499660912.2 (11.3 to 13.1)Ref186 0668 0534.3 (4.2 to 4.4)Ref2 587 499384 35114.9 (14.8 to 14.9)Ref
Implantable968 893206 48421.3 (21.2 to 21.4)<0.00184163954.7 (4.2 to 5.1)<0.001784 528189 99824.2 (24.1 to 24.3)<0.001175 94916 0919.1 (9.0 to 9.3)<0.001
Breakthrough status
Non-breakthrough3 743 589600 39616.0 (16.0 to 16.1)Ref13 3179867.4 (7.0 to 7.8)Ref969 253197 97820.4 (20.3 to 20.5)Ref2 761 019401 43214.5 (14.5 to 14.6)Ref
Breakthrough38651012.6 (2.1 to 3.1)<0.001951818.9 (11.1 to 26.8)<0.0011341735.4 (4.2 to 6.7)<0.0012429100.4 (0.2 to 0.7)<0.001
  • Report time defined based on difference between day Food and Drug Administration reported receiving device report and day manufacturer reported becoming aware of device event. Late reports were defined as those with report times greater than 30 days. Only reports with non-missing and non-negative report times included in statistical analyses. Statistical tests of differences in percentages compared percentages to reference group (denoted with Ref). Q values adjusted two sided P values for multiple testing and were estimated via the Simes procedure based on 64 total tests of statistical significance in analyses.

  • CI=Confidence interval

Table 4

Differences in percent reported late by event type and preauthorization device characteristics

Reports related to lifesaving devices and implantable devices had higher percentages of late reports compared with non-lifesaving and non-implantable devices for injury reports but lower percentages of late reports for death and malfunction reports (table 4).

Breakthrough devices had less late reporting compared with non-breakthrough devices: 2.6% of the 3865 reports for breakthrough devices were reported late (2.1% to 3.1%), compared with 16.0% of the 3 743 589 reports for non-breakthrough devices (16.0% to 16.1%, Q<0.001). Results were similar for injury and malfunction reports. However, among death reports, 18.9% (11.1% to 26.8%) were late for breakthrough devices, compared with 7.4% (7.0% to 7.8%, Q<0.001) for non-breakthrough devices.

Differences in mean report times were qualitatively similar for device class and implantable status, but lifesaving devices had a lower mean report time for injuries compared with non-lifesaving devices and breakthrough devices had a lower mean report time for deaths compared with that of non-breakthrough devices (appendix table 5).

Late reporting by recall status

For reports related to class II and class III devices without missing or invalid report times, associations between recall status and report times varied by the adverse event type. Overall, class II and III devices with ongoing recalls and terminated recalls (recalls since resolved as manufacturers completed necessary corrective and preventive actions) had less late reporting compared with non-recalled devices. Of the 1 277 990 reports for devices with at least one ongoing recall, 201 058 were late (15.7% (95% CI 15.7 to 15.8%)), compared with 394 573 of the 2 399 304 reports (16.4% (16.4% to 16.5%, Q<0.001)) for devices with no ongoing recalls (table 5). Similarly, 41 039 (13.9% (13.7% to 14.0%)) of the 296 061 reports for devices with at least one terminated recall were late, compared with 554 592 (16.4% (16.4% to 16.4%)) of the 3 381 233 reports for devices with no terminated recalls (Q<0.001).

Table 5

Differences in percent reported late by event type and recall status

Recall statusAllDeathInjuryMalfunction
Total No.Reported lateQTotal No.Reported late QTotal No.Reported lateQTotal No.Reported late Q
No.% (95% CI)No.% (95% CI)No.% (95% CI)No.% (95% CI)
Ongoing recall
No ongoing recall2 399 304394 57316.4 (16.4 to 16.5)Ref12 7908957.0 (6.6 to 7.4)Ref855 057172 00820.1 (20.0 to 20.2)Ref1 531 457221 67014.5 (14.4 to 14.5)Ref
At least one ongoing recall1 277 990201 05815.7 (15.7 to 15.8)<0.0015189718.7 (15.4 to 22.1)<0.001109 17724 95822.9 (22.6 to 23.1)<0.0011 168 295176 00315.1 (15.0 to 15.1)<0.001
Terminated recall
No terminated recall3 381 233554 59216.4 (16.4 to 16.4)Ref12 6309237.3 (6.9 to 7.8)Ref846 592170 46420.1 (20.0 to 20.2)Ref2 522 011383 20515.2 (15.2 to 15.2)Ref
At least one terminated recall296 06141 03913.9 (13.7 to 14.0)<0.0016786910.2 (7.9 to 12.5)0.007117 64226 50222.5 (22.3 to 22.8)<0.001177 74114 4688.1 (8.0 to 8.3)<0.001
  • Report time defined based on difference between day Food and Drug Administration reported receiving device report and day manufacturer reported becoming aware of device event. Late reports were defined as those with report times greater than 30 days. Only reports for class II and class III medical devices with non-missing and non-negative report times included in statistical analyses. Statistical tests of differences in percentages compared percentages to reference group (denoted with Ref). Q values adjusted two sided P values for multiple testing and were estimated via the Simes procedure based on 64 total tests of statistical significance in analyses.

  • CI=confidence interval

Table 5

Differences in percent reported late by event type and recall status

In contrast to the overall results, death and injury reports for recalled devices were more likely to be reported late than devices with no ongoing recalls: 97 (18.7% (95% CI 15.4% to 22.1%)) of 518 death reports for devices with ongoing recalls were late compared with 865 (7.0% (6.6% to 7.4%)) of 12,790 death reports for devices with no ongoing recalls (Q<0.001; table 5). 22.9% (22.6% to 23.1%) of the 109 177 injury reports for devices with ongoing recalls were late, compared with 20.1% (20.0% to 20.2%) of the 855 057 injury reports for devices with no ongoing recalls (Q<0.001; table 5). Similar patterns in late death and injury reporting were present for terminated recalls.

When looking at mean report times rather than late reporting rates, devices with ongoing recalls had a higher mean report time compared with devices with no ongoing recalls, while devices with terminated recalls had a lower mean report time compared with devices with no terminated recalls (appendix table 6).

Discussion

Principal findings

In this cross sectional analysis of over 4 000 000 medical device adverse event reports submitted by manufacturers to the FDA’s MAUDE database between 1 September 2019 and 31 December 2022, over 1 000 000 reports (nearly a third of all reports) were not demonstrably reported within the timeframe allowed by the FDA. Over 400 000 reports were reported more than six months (180 days) after manufacturer notification of the event, representing more than two thirds of all late reports. Many of the devices with large numbers of late reports were crucial to patient care, including devices like infusion pumps used in hospitals and continuous glucose monitors used by patients in ambulatory settings. Late reporting to the MAUDE database could preclude identification of patient safety concerns in a timely manner.

Policy implications

Late death reports are potentially the most serious omission from MAUDE and were concentrated in problematic areas. Breakthrough device deaths were reported late more often than non-breakthrough device deaths. This finding is concerning because breakthrough devices may have less developed premarket evidence supporting their safety and efficacy,124546 meaning any late reporting for breakthrough devices could have an outsized public health impact relative to late reporting for non-breakthrough devices. However, fewer than 100 death reports were made for breakthrough devices, and injuries and malfunctions were more often reported on time for breakthrough devices. Nonetheless, the program’s rapid growth in combination with the potential public health impact of any late reporting for breakthrough devices suggests greater attention is warranted regarding their safety and reporting practices.36

Class II and III (intermediate and high risk) recalled devices had a lower overall late reporting rate but higher late reporting rates for deaths and injuries compared with devices with no recalls. Put differently, recall initiation was followed by the reporting of serious adverse events that should already have been reported to the FDA, according to regulations. Additionally, late reports were disproportionately released in large batches and in short time periods rather than consistently over time. For example, 7% of late reports in the study were attributable to one glucose monitor (table 3). This device had a relatively constant flow of on-time reports during the sample period, but nearly 100% of its late reports were released in March 2021 (appendix fig 10), the same month the manufacturer issued a recall for the monitor.4748 This bunching pattern, observed both when looking at specific medical devices with large numbers of late reports and the entire sample (appendix fig 5-18), is consistent with media reports of manufacturers only reporting important safety information after recall initiation or other high profile events.1819 Releasing large batches of late reports may be especially concerning, as clinicians or patients may have been able to respond to the earliest reports of deaths and injuries within the batch and avoid later ones had the reports been released more consistently.

Reporting delays could stem from manufacturers knowingly withholding important safety information from the public, as has been previously reported.1819202122 At the same time, delays could also represent the time required for manufacturers to verify adverse events and gather additional information before reporting events to the FDA. Nonetheless, late reporting is not permitted under existing regulations, regardless of intent. Withholding safety information may cause avoidable patient harm given the role the MAUDE database currently has in identifying emerging safety issues.13

The findings from this study collectively show that while the MAUDE database often informs FDA safety actions,13 this data source is incomplete for understanding medical device safety issues due to late adverse event reporting from manufacturers. Besides impacting how policy makers, clinicians, and patients make medical decisions, this may affect future device development. In a draft guidance for its largest device authorization pathway, the FDA has proposed that manufacturers use MAUDE reports to establish safety profiles for devices under review.49 As these reports may be missing, manufacturers may incorrectly conclude that devices are safe by interpreting a lack of reports as a lack of adverse events rather than improper reporting.

Policy actions could address late reporting. Most reports were not late, and most late reports were submitted by just a few manufacturers, indicating that on time reporting is generally feasible under existing regulations. Accordingly, a limited number of FDA warning letters and follow-up enforcement actions related to consistently violative manufacturers may sufficiently deter late reporting. Mechanisms for enforcing manufacturer reporting requirements include “seizure, injunction, civil money penalties, and criminal prosecution.”50 However, the FDA instead typically “relies on the goodwill and cooperation” of manufacturers, rather than using the full set of sanctions available to the agency.51 Insufficient enforcement may give manufacturers the impression that late reporting will be tolerated and result in ultimately harming patients by delaying access to important safety information.

Apart from stronger enforcement, the FDA and other organizations have intermediate steps they could pursue. The FDA could require manufacturer notification dates on electronic submissions of reports.52 Given that almost all reports with invalid date information simply did not report the manufacturer notification date, this change would likely improve the quality of reports in MAUDE. Additionally, the FDA could consider regularly publishing a list of manufacturers or devices associated with large numbers of late reports. Such a list could help to inform patient and clinician decision making regarding medical device selection, as well as spur manufacturers to submit timely reports for the sake of avoiding negative media attention. The FDA has previously been reluctant to pursue strategies that involve singling out violative manufacturers.53 However, other government and non-government organizations could publish a similar list, given that the MAUDE database is publicly accessible.

More generally, this study highlights the limitations of passive postmarket surveillance systems that rely on manufacturers and other non-government entities to report safety information to regulators, as well as the need to develop more active postmarket surveillance mechanisms. The US is not unique in this regard, as many countries primarily use passive surveillance mechanisms to detect emerging device safety issues.54 Efforts to routinely collect standardized information on medical device usage and outcomes in national and global databases, such as integrating unique device identifiers into administrative claims and electronic health records, would likely provide a more comprehensive assessment of real-world safety outcomes compared with relying on manufacturers and others to passively report information.555657 Such efforts would also complement existing surveillance mechanisms such as device registries and postmarket studies ordered by the FDA, which are only available for a limited number of devices.5859

Efforts to implement safety surveillance systems based on unique device identifiers would require few additional resources from medical device manufacturers, as most manufacturers already comply with the FDA’s 2013 final rule to include unique device identifiers on all medical devices.60 Instead, the costs of implementing such systems would likely fall on federal regulators and health care providers. The FDA estimates that establishing and maintaining an active medical device postmarket surveillance system would cost the agency US$8 million per year.61 Health care providers also often face logistical difficulties in initially implementing safety surveillance system integration into medical records, but those that do report clinical and operational benefits as well as revenue potential.6062636465666768 In short, while establishing an active medical device surveillance system is not without its costs and challenges, doing so is feasible and would represent an important advance relative to the limitations of passive surveillance approaches as identified in this study.

Limitations

This study had several limitations. Firstly, measuring report times depended on dates reported by manufacturers, which may have been misreported. Secondly, MAUDE does not differentiate between reports with a five work day reporting requirement versus a 30 day requirement. This study only characterized report times greater than 30 days as late, meaning late reporting may be underestimated. However, an older analysis conducted by the FDA show that less than 0.1% of adverse event reports were subject to the five work day reporting requirement as of 2007, potentially indicating that the extent to which our study underestimates late reporting may be small.69 Thirdly, MAUDE cannot identify unreported adverse events, including events where manufacturers were not notified and events where manufacturers were notified but did not report.

Fourthly, this study did not directly identify harms caused by late reporting. Identifying such harms is difficult without a nationwide active surveillance infrastructure, as the relevant exposures (late reporting in MAUDE) and safety outcomes (for which MAUDE would be the likely data source) are necessarily linked. Fifthly, analyses examining associations between device characteristics and report times were exploratory and not causal. This study was unable to determine why manufacturers report late and their underlying motivations for doing so. However, late reporting is not permitted under existing regulations, meaning greater policy attention is warranted regardless of the cause.

Conclusions

This study found that during a recent three and a half year period, nearly a third of initial manufacturer reports of adverse events in the MAUDE database were not demonstrably reported on time, with over two thirds of late reports submitted more than six months after manufacturer notification of events. More than 50% of late reports were attributable to three manufacturers. MAUDE is an incomplete data source for understanding medical device safety issues, due partly to the timeliness of its reports.

What is already known on this topic

  • Medical device adverse event reports in the US Food and Drug Administration’s primary postmarket surveillance database have known concerns about their reliability

  • Multiple media reports suggest that adverse events reports from medical device manufacturers may not be submitted to the FDA within the required 30 day deadline set by federal regulations

What this study adds

  • Over 1.2 million medical device adverse event reports (nearly a third of the study sample) were not demonstrably submitted to the FDA within the deadline set by federal regulations

  • More than half of late medical device adverse event reports were submitted by three manufacturers

  • Patient safety concerns may not be identified in a timely manner due in part to late manufacturer reporting of medical device adverse events

Ethics statements

Ethical approval

This study used publicly available materials and did not involve humans, therefore, ethics committee approval was not required.

Data availability statement

Statistical code and datasets are available on Mendeley Data at https://data.mendeley.com/datasets/mydr3vdzcr/1.

Footnotes

  • Contributors: All authors contributed to the design and conduct of the study; management, analysis, and interpretation of the data; and preparation, review, or approval of the manuscript. AOE is the guarantor. The corresponding author attests that all listed authors meet authorship criteria and that no others meeting the criteria have been omitted.

  • Funding: None.

  • Competing interests: All authors have completed the ICMJE uniform disclosure form at www.icmje.org/disclosure-of-interest/ and declare: AOE reports prior employment by Medtronic, personal fees from the Digital Medicine Society and the University of Southern California, and grant funding from the Agency for Healthcare Research and Quality, the National Institute on Ageing, and the National Bureau of Economic Research. PK-M reports unrelated research funding from the American Cancer Society, Robert Wood Johnson Foundation, Agency for Healthcare Research and Quality; unrelated consulting services to Koya Medical and Mayo Clinic; and unrelated role serving in executive position and holding equity in XanthosHealth, a company developing a social care referral platform. RFR reports research funding from Arnold Ventures and serves as a member of the Institute for Clinical and Economic Review California Technology Assessment Forum. JSR receives research support through Yale University from Johnson and Johnson to develop methods of clinical trial data sharing, from the Food and Drug Administration for the Yale-Mayo Clinic Center for Excellence in Regulatory Science and Innovation program (U01FD005938), from the Agency for Healthcare Research and Quality (R01HS022882), and from Arnold Ventures; formerly received research support from the Medical Device Innovation Consortium as part of the National Evaluation System for Health Technology and from the National Heart, Lung and Blood Institute of the National Institutes of Health (R01HS025164, R01HL144644); JSR was also an expert witness at the request of Relator's attorneys, the Greene Law Firm, in a qui tam suit alleging violations of the False Claims Act and Anti-Kickback Statute against Biogen Inc that was settled September 2022. SSD reports research funding from the Department of Veterans Affairs and Arnold Ventures, serving as a member of the Institute for Clinical and Economic Review California Technology Assessment Forum, and serving on the Medicare Evidence Development and Coverage Advisory Committee. All other authors declare no competing interests.

  • Transparency: The lead author (the manuscript’s guarantor) affirms that the manuscript is an honest, accurate, and transparent account of the study being reported; that no important aspects of the study have been omitted; and that any discrepancies from the study as planned (and, if relevant, registered) have been explained.

  • Dissemination to participants and related patient and public communities: Results will be shared via the dissemination teams at Washington University in St Louis, University of Minnesota, University of California, San Francisco, and Yale University. Typical mediums include press releases, social media posts (X, Instagram, Facebook, and LinkedIn), and emails sent directly to journalists representing outlets such as the New York Times, STAT, ProPublica, Regulatory Focus, and Becker’s Hospital Review, which summarize the study’s main findings. The authors will publish a plain language summary of the study and its implications to a publicly available website. The authors will also provide public comments to government agencies when agencies solicit public input.

  • Provenance and peer review: Not commissioned; externally peer reviewed.

http://creativecommons.org/licenses/by-nc/4.0/

This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/.